1. Introduction
Test Standards
→GB/T50081 Ordinary concrete mechanics performance test method standard
→GB28635 Concrete paving bricks
→GB11968 Autoclaved aerated concrete blocks
→GB 13544 Fired perforated bricks
→GB/T 2542 Test method for wall bricks
→GB/T5101 Fired common bricks
→GB/T8488 Acid-proof bricks
→GB/T4111 Test methods for the small concrete hollow block
→GB13545 Fired hollow bricks and blocks
→GB/T15229 light aggregate concrete small hollow blocks
2. Technical Speicfication
YAW-500C/600C/800C/1000C.500kN/600kN/800kN/1000kN |
||
1 |
Max. Load capacity |
500kN/600kN/800kN/1000kN |
2 |
Relative error of indicating value of load |
≤±1.0% |
3 |
Measuring range of load |
2%-100%F.S |
4 |
Load resolution |
1/±300000F.S |
5 |
Control range of constant speed stress |
2-60 N/mm2·S-1 |
6 |
Stress speed error |
≤±5% |
7 |
Control range of constant speed strain |
0.00025/s-0.0025/s |
8 |
Strain speed error |
≤±5% |
9 |
Displacement resolution |
0.01mm |
10 |
Loading speed |
0-60kN/s |
11 |
Piston stroke |
≥150mm |
12 |
Displacement speed of piston |
≥150mm/min |
13 |
Distance between upper and lower platens |
0-450m |
14 |
Moving speed of cross head |
≥150mm/min |
15 |
Size of upper pressure platen |
ø150mm;ø200mm;ø300mm;400x400mm/optional |
16 |
Size of lower pressure platen |
400x400mm |
17 |
Span between columns(L*W) |
500mm |
18 |
Working voltage |
380V±10% 50Hz |
19 |
Power |
1.5kw |
20 |
Dimension of loading frame(L*W*H) |
Approx. 980x1080x1800mm |