→Determine the impact resistance performance of metallic materials under dynamic loads using Charpy impact test (V‑notch and U‑notch specimens), so as to evaluate the dynamic‑load properties of materials.
→Automatic specimen temperature control: specimen cooling, constant‑temperature holding, user‑defined parameter setting, and fully‑automatic test execution.
→Supports continuous impact test and batch impact test for rapid, accurate and efficient operation.
→The computer automatically collects, processes and stores impact absorbed energy, impact toughness, pendulum lifting angle and average test values.
→Brake system and power‑failure pendulum‑holding device prevent accidental pendulum release for high safety and reliability.
→Fully‑enclosed safety protective guard, constructed of aluminum alloy profiles and organic glass, featuring elegant appearance and reliable safety protection.
→Electronic safety‑door design: when the protective‑guard hatch door is opened, the equipment immediately stops the pendulum movement and firmly locks the pendulum.
→Equipped with a specimen‑recovery device, which operates automatically upon completion of each impact.
→Compliant with GB/T 229 Metallic materials‑Charpy pendulum impact test method;
→GB/T 3808 Verification of pendulum‑type impact testing machines;
→ISO 148‑1 Metallic materials — Charpy pendulum impact test — Part 1: Test method;
→ISO 148‑2 Metallic materials — Charpy pendulum impact test — Part 2: Verification of testing machines;
1. Impact energy: 450 J
2. Pre‑lifting angle of pendulum: 150°
3. Distance from pendulum shaft center to impact point: 800 mm
4. Impact velocity: 5.2 m/s
5. Span of specimen supports: 40 mm
6. Jaw fillet radius: R1‑1.5 mm
7. Impact blade fillet radius: R2‑2.5 mm
8. Thickness of impact blade: 16 mm
9. Temperature‑control range of low‑temperature bath: ambient temperature ~ −80 ℃
10. Automatic specimen‑feeding impact cycle time: 4.5 s